LeCroy Oscilloscope
LeCroy
CHAUVIN ARNOUX(샤빈 아녹스)
IWATSU
Coherent Solutions
ODA(Power/ Load)
제전기 ( IONIZER )
Magna Power
NHR
APT
JENOPTIK
GWINSTEK
FLUKE
YOKOGAWA
AMETEK
NOISETECH
열충격시험기
Spellman
▶SL
▶205B
▶210
▶SL150KV
▶SLS
▶SA
▶ST
Spellman > 210
PRODUCTS Spellman > 210
210 Series Linear Rack Mount Supply
기계적 구성
파워
출력 전압
Rack Mount
125W, 135W, 140W, 150W, 195W, 200W, 225W
Model 210 is available in 9 output voltages ranging from 1kV to 50kV
Spellman’s Bertan brand of 210 Series of 125 to 225 watt
high voltage power supplies provide regulated high voltage
outputs from 1 to 50kV. The low noise, linear topology
employed results in extremely low output ripple specifications.
Units are inherently reversible by design, providing
either positive or negative output polarity. The 210 is
fully arc and short circuit protected. Excellent regulation
specifications are featured along with outstanding stability
performance.
TYPICAL APPLICATIONS
HiPot Testing
CRT Testing
Electrostatics
E Beam Systems
General Laboratory Usage
SPECIFICATIONS
Input Voltage:
115Vac, ±10%, 50/60 Hertz @ 5 amps
230Vac, ±10%, 50/60 Hertz @ 2.5 amps
Input voltage is switch selectable
Output Voltage:
See “model selection” table
Output Polarity:
1kV to 50kV units are inherently reversible by design
Output Current:
See “model selection” table
Voltage Regulation:
Line: ≤0.001% of rated output voltage over specified
input voltage range
Load:≤0.005% of rated output voltage for a full load
change
Current Regulation:
Internally set to limit at 105% of rated current at full
output voltage. Maximum output current at any other
voltage setting must be derated linearly down to 30%
of maximum at zero output voltage
Ripple:
See “model selection” table
Temperature Coefficient:
≤50ppm/°C
Stability:
≤0.01%/hour, 0.02% per 8 hours after a 1/2 hour warm up
Accuracy:
Voltage Monitor: ±(0.25% of reading + 0.25% of maximum)
Current Monitor: ±(0.5% of reading + 0.25% of maximum)
Remote Programming; ±(0.25% of setting + 0.05% of
maximum) for 1kV to 30kV ±(0.5% of setting + 0.25% of
maximum) for 50kV
Front Panel Control: ±(0.25% of setting + 0.05% of maximum)
for 1kV to 30kV ±(0.5% of setting + 0.25% of maximum)
for 50kV
Front Panel Meter: ±2% of full scale
Operating Temperature
0°C to +50°C
Storage Temperature:
-40°C to +85°C
Humidity:
20% to 85% RH, non-condensing
Input Line Connector:
A captive 3 conductor line cord and NEMA plug is provided
Interface Connector:
7 pin Amphenol 126-198, mating connecter and pins provided
Output Connector:
A detachable 10 foot (3 meter) long HV cable is provided
Cooling:
Internal fan, forced-air cooling
Dimensions
1-5kV: 19.0˝ W X 5.25˝ H X 11.0˝ D
(483mm X 133mm X 279mm)
10-50kV: 19.0˝ W X 5.25˝ H X 16.0˝ D
(483mm X 133mm X 406mm)
Weight:
≤40 pounds (18.1kg) up to and including 30kV units
≤50 pounds (22.7kg) for 50kV unit
Regulatory Approvals:
Compliant to 2004/108/EC, the EMC Directive
and 2006/95/EC, the Low Voltage Directive.